Take five different concentrations of organic solvent comparison under test ( Linear range according to the actual sample under test of organic solvent content determine) 。 At 100 ℃ & plusmn; 2 ℃ constant temperature box maintain 60 minutes ( When necessary, according to material and solvent, under test is placed within the range 80 ~ 150 ℃ for 15 minutes and 60 minutes, decomposition temperature is higher than the solvent boiling point below materials as appropriate) With preheating temperature to the same container the top 1 ml syringe extraction of gas into the gas chromatograph. Measurement of peak area ( Or peak) , draw peak area ( Peak) The standard curve with the corresponding organic solvent quality. Take the sample, according to the standard curve method 'at 100 ℃ & plusmn; 2 ℃ constant temperature box keep 60 minutes' operation. According to the sample under test in the organic solvent peak area ( Peak) , from the standard curve of the quality of the samples contained in the solvent, press type calculation. X = m/W type: X-ray - - Samples of solvent residue, mg/m2; m - - - To test the quality of the solvent, mg; W - - - Sample surface area/m2.